Rice Lake Explorer Series Toploader, Ohaus User Manual

Page 63

Advertising
background image

EXPLORER BALANCES

EN-61

4.10 Density

Determination

The Explorer can be used to determine an object’s density. Four types of density determination can be made:

1. Solids

more dense than water

2. Solids

less dense than water

3. Liquid

density

4. Porous material (impregnated with oil)

A density determination kit is designed to be used with Ohaus Explorer balances. Illustrations in this procedure
refer to the density kit. However, you may use whatever lab apparatus that suits the requirements for density
measurements. The balance software includes a built-in reference density table for water at temperatures
between 10°C and 30°C. Review this entire section before attempting density measurements.

When making density measurements the material should weigh at least 10 mg on an analytical balance and 100
mg on a precision balance.

Density can be determined manually (with a key press after each step) or automatically (weights are
automatically recorded). Before making density measurements, establish the Application Settings.

The DENSITY DETERMINATION Home screen

Main Display Line
Second Display Line (not shown)

Reference Fields

Application Buttons

Functions Application Icon


4.10.1 Measuring the Density of a Solid Using Water (default)

Press the Setup button (the wrench) to open the Density
Determination Settings screen.

Confirm the following Setups are selected:

9

Density Type: Solid

9

Liquid Type: Water

9

Porous Material: Off

Press Done to return to the Density Determination home screen.

The Edit Settings Screen.

Settings available: Water Temp
Functions Available: Return to Application

To adjust the water temperature value,
touch the Water Temp button.
The balance calculates water density based on the water
temperature value entered (internal lookup table).
Measure the actual water temperature using a precision
thermometer.

Advertising