5.5 Balance

Setup

Enter this sub-menu to customize Balance display functionality.

5.5.1 Balance Set-up sub-menu

Stability

Indicator Range

Filter Level

Auto Zero

Tracking

Auto Tare

Gross Indicator

Legal For Trade

Graduations

Date & Time


Factory default settings are shown below in bold.

5.5.2 Stable

Range

Set the amount the reading can vary while the stability symbol remains on.

0.5 D

= 0.5 graduations

1 D

= 1 graduation

2 D

= 2 graduations

3 D

= 3 graduations

5 D

= 5 graduations

Note: When LEGAL FOR TRADE (below) is set to ON, the setting is forced to 1 D.
The setting is locked when the Security Switch is set to the ON position.

5.5.3 Filter

level

Set the amount of signal filtering.

LOW

= faster stabilization time with less stability.

MEDIUM = normal stabilization time with normal stability.
HIGH

= slower stabilization time with more stability.

5.5.4 Auto Zero Tracking
Set the automatic zero tracking functionality.

OFF =

disabled.

0.5 D

= display maintains zero up to a drift of 0.5 graduation

per second

1 D

= display maintains zero up to a drift of 1 graduation per

second.

3 D

= display maintains zero up to a drift of 3 graduations per

second.


Note:
When Legal for Trade is set to ON, the AZT setting is forced to 0.5 D.
The OFF setting is still available. The setting is locked at the current setting when the
security switch is set to the ON position