Rice Lake Explorer Series Toploader, Ohaus User Manual
Page 81
EXPLORER BALANCES
EN-79
4.13.2 Begin Pipette Adjustment Process
Touch
Begin Pipette
Adjustment.
Follow the
screen
instructions, to
Place
container on
the pan. Then
Press Tare.
Step 1 – Dispense the First Sample
Follow the screen instructions, then press Accept to store the
first sample weight (liquid).
Continue to follow the screen instructions by dispensing
samples and pressing Accept to store each sample weight
(liquid).
The default number of samples is 10.
It can be changed in Setup, from 2 to 30 samples.
Once all the sample
weights have been
dispensed,
the Results screen is
automatically
displayed.
Press Graph to see
results plotted. Press
Result to return to the
result screen. Press
Close to return to the
Pipette Adjustment
home screen.
Note: To view either the results or graph screens again from the home screen, press the View Result button.
To start a new Pipette Adjustment process, press Begin Pipette Adjustment.