Rice Lake Explorer Series Toploader, Ohaus User Manual

Page 91

Advertising
background image

EXPLORER BALANCES

EN-89

5.2 Main

Menu

The Explorer balance Main menu structure is illustrated below.

Calibration User

Settings

Balance

Setup

Application

Modes

Weighing

Units

GLP and

GMP Data

Communication

Library

I/O

Settings

Diagnostics

Factory

Reset

Lockout Information

5.3 Calibration

Explorer Balances offer a choice of six calibration methods: Internal Calibration, AutoCal Calibration,
Calibration Adjust, Span Calibration, linearity Calibration, and User Calibration.

Note: Do not disturb the balance during any calibration.

5.3.1 Calibration

sub-menu

AutoCal™

Internal

Calibration

Automatic

Calibration

AutoCal™

Adjustment

Span

Calibration

User

Calibration

Calibration

Test


5.3.2 Internal Calibration (AutoCal™)
Calibration is accomplished with the internal calibration mass. Internal calibration can be performed at
any time, provided the balance has warmed up to operating temperature and is level.

With the Balance turned ON and no load on the pan, touch AutoCalInternal Calibration.
The Balance begins to calibrate.

Alternately, touch the shortcut Calibration Icon from any Application home screen
to initiate an Internal Calibration.

The display shows the status, then returns to the current application.

To cancel at any time, press Cancel.

5.3.3 Automatic

Calibration

When Automatic Calibration is set ON, the balance performs a self calibration when a measured
predefined change occurs (in temperature or time). AutoCal will automatically calibrate the Balance (using
the internal mass) each time a change in temperature significant enough to affect accuracy, or after every
11 hours.