EN-26

EXPLORER

BALANCES

4.2.3 Parts Counting – Check

This feature permits establishing a set quantity of pieces as criteria for similar items that can be quickly checked
against a sample.

Press the
Setup
(wrench)
button to enter
the Counting
Setup menu.

Change
Counting
Mode to
Check
Counting
.

The PARTS COUNTING – Check Home screen

Main Display Line

Second Display Line

Reference Fields

Application Buttons

Functions

The default (or last) Average Piece Weight (APW) is
displayed. Place objects on the pan to display the
Check Status (number of pieces).

4.2.2.1 Edit Settings: To use the Parts Counting-Check mode,
the Over Limit and Under Limit values must be set.

The Average Piece Weight and standard sample size can also
be set. Follow these steps:

Touch the Edit Settings button. The Settings screen appears.
The buttons for APW, Samples, Over Limit and Under Limit
show default (or previously set) values. To change any value,
touch the button. A keyboard screen appears.
(See keyboard screens, next page.)

To set a new average piece weight, press the
Recalculate APW button.

A new screen appears with the message:

Place 30 pieces. Press Accept
(Default is 10 pieces, but message matches
number of pieces in Samples button.)

After following these instructions, press Return to Application
on the main Edit Settings screen.

Advertising