EXPLORER BALANCES

EN-91

5.3.6 User

Calibration

User Calibration is used to calibrate the balance using a user defined mass.

With no load on the scale, press User Calibration from the calibration sub-menu.
The display shows the current full load (span) calibration point and calibration unit of measure.

To change the value to a user-defined weight, touch ENTER VALUE.
Enter the weight via the popup screen. The balance then prompts for the User-defined weight.

Place the specified calibration weight on the balance and the calibration process completes.
The display shows status, then returns to the current application.

5.3.7 Calibration

Test

Use Calibration Test to compare a known calibration weight against the stored span calibration data.

With no load on the scale, press Cal Test to begin the process.
The balance captures the zero point, then prompts for the span value.

The display shows status, followed by the difference between the current calibration weight and the
stored calibration data.

After 3 seconds, the display returns to the current application.