5.7.2 Turning a Unit ON/OFF

Select the desired unit, then touch the item to
toggle it either on or off.

The current menu item status is shown.

OFF =

disabled

ON =

enabled


5.7.3 Custom

Units

Use each Custom Unit to display weight in an alternative unit of measure. The custom unit is defined using a
conversion factor, where the conversion factor is the number of custom units per kilogram expressed in scientific
notation (Factor x 10^Exponent).

For example: To display weight in troy ounces (32.15075 troy ounces per kilogram)
enter a Factor of 0.321508 and an Exponent of 2.

Set the status.

OFF =

disabled

ON =

enabled


When Custom Unit is set to ON, the Factor, Exponent and Least Significant Digit must be set.

Factor
Set the conversion factor using the numeric keypad.

Settings of 0.00001 to 1.9999999 are available. The default setting is 1.000000


Exponent
Set the factor multiplier.

0

= multiply the Factor by 1 (1x10

0

)

1

= multiply the Factor by 10 (1x10

1

)

2

= multiply the Factor by 100 (1x10

2

)

3

= multiply the Factor by 1000 (1x10

3

)


–3

= divide the Factor by 1000 (1x10

-3

)

–2

= divide the Factor by 100 (1x10

-2

)

–1

= divide the Factor by 10 (1x10

-1

)


Least Significant Digit
Set the graduation.

Settings of 0.00001, 0.00002, 0.00005, 0.0001, 0.0002, 0.0005, 0.001, 0.002,
0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500 and 1000
are available.


Note: Least Significant Digit setting selections are dependent on the Factor and Exponent settings.