EN-44

EXPLORER

BALANCES

4.5.5 Statistics for Dynamic Weighing

Use the Statistics function to compare a number of samples and examine the relative deviation of the samples
along with other statistical data. A minimum of three samples is required.
Statistics can be determined manually (Manual mode, with key press after each step) or automatically
(Automatic Mode: weights are automatically recorded when stable). To use the statistics function, enable it first
in the Application Setup menu.


Application Setup

First, Enable Statistics

Touch the Setup
Icon wrench to
enter Setup
menu.

Press Statistics.
Choose either
Manual or
Automatic
mode, then
press Done.


Accumulating Displayed Values & Viewing Data – Manual
Once Statistics is enabled, two buttons appear on the Application screen: Statistics and Accumulate.

To add the sample to the data set, place the item on the pan
and press Accumulate.

When stable, the Statistics button momentarily lights to show
the item is added to the set and the data set quantity increments
by one.

Continue adding items and pressing Accumulate to build the
statistical data set

At any time, press the Statistics

button to view the results.

The Statistics Results screen

Press Print to send to the enabled COM port
Press Close to return to the Application home screen
with the current statistics.

Press Clear Data to clear the results and return to the
Application home screen.

Statistics – Automatic
Statistics Automatic mode is similar to manual mode (above) except the first stable weight is automatically
captured. Thus the Accumulate button is not used.

Advertising