Zero – Rice Lake MSI4300 PortaWeigh Plus Crane Scale 7 Key User Manual

Page 10

Advertising
background image

MSI-4300 Porta-Weigh Plus/7 Key

User Guide Page 9

M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L

2)

Display Check: All segments and the Total LED are illumi-
nated for 1 second.

3)

Message displays “MSI PAW+”, numerics give the software
version number (1 second).

4)

During display test and MSI message, all internal operations
are checked and any nonconformance causes an error
message to be displayed.

Final:
Unit reads the current weight in the last set mode (NET, GROSS,
VIEW TOTAL, PEAK NET, PEAK GROSS).

ZERO

Function:
Sets the zero reading of the scale.

Use:

1)

The scale must be stable. The scale will not zero if the
motion detect annunciator is on.

2)

The scale must be in the Gross mode. The scale will not zero
in the Net or Total modes.

3)

The scale will accept a zero setting over the full Range of the
scale (Legal-for-trade models might have a limited zero
range). Zero settings will subtract from the overall capacity
of the scale. For example, if you zero out 1000 lbs on a 5000
lb scale the overall capacity of the scale will reduce to 4000
lbs plus the allowed overrange amount.

Action:
Push ZERO button.
The weight reading must be stable within

±

1 division. If this condition

is met the display reads “ZEROED” and the digits display 0 (or 0.0 or
0.00 depending on resolution). The zero setting is stored by the
backup memory, which will restore it the next time the system is
turned on.

Advertising