Rice Lake MSI4300 PortaWeigh Plus Crane Scale 7 Key User Manual

Page 36

Advertising
background image

MSI-4300 Porta-Weigh Plus/7 Key

User Guide Page 35

M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L

Baud Rate:

300, 600, 1200, 2400, 4800, 9600, 19200, 38400

Data Bits:

7 or 8

Parity:

None, Even, Odd

Start Bits:

1

Stop Bits:

1 or 2

Data Format:

Each transmitted reading consists of 24 to 33 characters including 10
customizable characters. The preamble (five characters) is intended

F

Post String

A

Preamble

# 1

sp

X X X X X X X X

sp

Y Y Y Y Y Y Y

sp

Z Z Z

Data Field

B

C

D

E

for

printer control characters but can be any ASCII characters. The post
string (five characters) is used for Line Feed or other end-of -line
control characters. In addition, there is a 5 character initialization
string transmitted when the Porta-Weigh Plus is turned on. These three
character strings are custom programmed to customer specifications
into the Porta-Weigh Plus at the factory. The standard configuration is
optimized for a 24 column printer, available from several manufactur-
ers. Contact the factory for a list of plug compatible printers. The
transmitted data is grouped in fields, A through F, defined as follows:

A. Preamble field: up to 5 characters can be sent for control

characters and printer formatting. The standard configuration
eliminates the preamble.

B. ID Code Field: the first character is the number sign, the

second character is a digit from 1 to 9 indicating the ID
number.

C. Weight Field: eight characters of weight or total data. “0” to

“9”, including a minus sign and a decimal point if appropri-
ate.

D. Mode Field: seven characters that indicate the printed weight

modes: GROSS, NET, TOTAL, PKGROSS, PKNET

E.

Units Field: three characters that indicate the weight unit: lbs,

Advertising