MSI-4300 Porta-Weigh Plus/7 Key

User Guide Page 23

M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L

setpoint actually operating. To enable the setpoint, push
SELECT until either the "GROSS“ mode or the "NET/
GROSS" mode is selected. In the “GROSS” mode the
setpoint will operate at the value set based on Gross Weight
regardless of any tare value being subtracted. If this is the
desired set point operation mode push ENTER. If the set
point should operate relative to a tared weight the
“NETGROSS” mode should be used. Use the SELECT key
to change from “GROSS” to “NETGROSS” and then push
ENTER. In this mode the set point will operate at the value
which represents either a Gross weight or a Net weight.

7)

The next selection is the Set Point “RESPONSE”...what
should the scale do when the Set Point value is reached? Push
ENTER. There are now three selections: The “NORMAL”
response turns on the internal MOSFET which is connected
to a user supplied alert device such as a siren or strobe light.
Use the SELECT key to change the response to
“BLANKLCD”. In this mode the liquid crystal display
weight reading blanks out and the message displays HI
SETPT. The MOSFET functions as before. Use the SELECT
key again to change to the “BLINKLCD” response. Weight
readings continue, but the display alternates between a
standard weight reading, and a “HI SET PT” indication. In all
three modes, the overload condition will override the display
mode.

8)

After the desired Set Point response is selected, push EN-
TER.

9)

The Set Point Menu is now back on the screen. The SELECT
key can be used to return to any of above operations
(VALUE, MODE, RESPONSE) if modifications need to be
made. To quit and return to normal scale operation use
SELECT until the message “EXIT?” appears and push
ENTER, or push CLEAR (no shift required). This returns
you to the Setup Menu. Again, push CLEAR to get out of the