Eia-232-d serial output option – Rice Lake MSI4300 PortaWeigh Plus Crane Scale 7 Key User Manual

Page 35

Advertising
background image

Page 34

MSI-4300 Porta-Weigh Plus/7 Key

User Guide

M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L

When the desired mode "ENABLED", or "DISABLED" is
displayed push ENTER.

Caution: Disabling Auto Zero Maintenance will degrade temperature
and drift performance of the Porta-Weigh Plus.
Disabling the AZM is only intended for certification testing.
Under no other circumstances should the scale be used with the AZM
disabled.

When calibration is finished:
Seal the Calibration Port.

Calibration cycle:
Measurement Systems International recommends that the Porta-Weigh
Plus is calibrated at least once a year by a qualified scale technician
with certified calibration standards.

EIA-232-D SERIAL OUTPUT OPTION

Function:
The EIA-232 output is used in conjunction with the Infrared Remote
Control to output weight and total data to a printer or can be used for
two-way communications with a computer.

Electrical Conformance:
The electrical characteristics of the serial output is configured to
conform to the EIA Standard EIA-232-D. Cable connections include
RXD, TXD, Ground, CTS, and RTS. (CTS / RTS handshaking is not
currently supported, consult factory)

Configuration:
The Porta-Weigh Plus serial port option is factory configured only.
Standard data configuration is:

Baud Rate:

9600

Data Bits:

8

Parity:

None

Start Bits:

1

Stop Bits:

1

By customer order, the standard configuration can easily be changed.
Available configurations include any combination of the following
settings: