Page 16

MSI-4300 Porta-Weigh Plus/7 Key

User Guide

M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L

TEST

METRIC

AUTO

REMOTE

0

MOTION

+

-

Function:
Provides a functional system test, and an on-demand display check
without disturbing the current weighment. Also provides calibration
verification in the form of a load cell Calibration number.

Action:
To start the test sequence: Press SHIFT & POWER.

1)

All digits (7 segment and 16 segment) count once from 0 to
9. All annunciators are tested including lb, kg, t, and the
Total LED.

2)

Battery condition is displayed next in the form: % of Battery
life with a figure from 1 to 100 on the digits. The % battery
life will vary depending on whether the backlight is on or off.

3)

The CAL number is displayed next. To ensure that the CAL
number is accurate the scale must be unloaded and in the
same units that it was calibrated in. The original CAL
number is printed on the right battery tube calibration label.
Cal number shifts of up to 10 counts are insignificant.

4)

Internal tests are performed to further ensure scale integrity.

Final:
Either the reading returns to the last condition, or an error message is
displayed.

SETUP

Function:
Allows Front Panel entry of seldom set parameters.