0 user defined function keys, User defined function keys – Rice Lake MSI-4260 Port-A-Weigh User Manual

Page 22

Advertising
background image

18 MSI-4260 Port-A-Weigh Operator’s Manual

3.0 User Defined Function Keys

The following function descriptions are for optional user defined functions that are

programmed on the front panel USER key or the two function keys (F1 & F2) on the

RF Remote Control. The functions TOTAL, VIEW TOTAL, and NET/GROSS are

available full time on the RF Remote Control. To enable any of the USER key

functions, you must set up the USER keys following the procedures in Section 4.0.

TEST

The TEST function provides an LED test that lights all LEDs at once, next the SW

Model number followed by the Software Version number, the battery voltage and then

performs a display test that counts from 00000 to 99999. Other internal tests are

performed and if any test fails, an Error Code will display. See Section 8.1 for a

troubleshooting guide. Press the

USER

key again within two seconds to enable a

single step through all the test functions. In the single step mode, use the

USER

key

to scroll through the available test functions and the

TARE

key to start or display the

individual tests. Use the

ZERO

key to exit individual tests. Use the

ZERO

key to exit

entirely from the TEST function.



MSI-4260 Software Version Number: Press

TARE

to see the Software

Version.



– Battery Voltage: Press

TARE

to see the battery voltage.



– Display Test: Press

TARE

to have the display count up from 00000 to

99999.



– Cal Constant Display: Press

TARE

to view the C-CAL value.

TOTAL

Note

The Total Mode must be programmed from the Setup Menus before the
USER key will function.

For accumulation of multiple weighments. The accumulator always uses the

displayed weight, so GROSS and NET readings can be added into the same TOTAL.

There are four Modes of totalizing: Manual and three Auto Modes. The Manual Mode

requires the TOTAL button be pressed with the weight on the scale. The weight will

be added to the previously accumulated value. This assures that a weight on the scale

is only added to the total once. Both the manual and three auto total modes require

that the weight on the scale return below 0.5% (relative to full scale) of GROSS

ZERO or NET ZERO before the next weighment can be added. Applied weight must

be ≥1% of full scale above GROSS ZERO or NET ZERO before it can be totaled.

MANUAL TOTAL

The

USER

key under the MANUAL TOTAL mode functions in this manner:

Weight is > 1% of Capacity and has not been totaled – Pushing the

USER

key will

add the current weight to the TOTAL weight. The ACK LED’s blink to indicate the

weight was accepted. The TOTAL annunciator lights and the Total weight is

displayed for five seconds and then the number of samples is displayed for two

seconds.
Current Weight has been Totaled – Pushing the

USER

key will display the Total

weight for five seconds (View Total) without changing the Total value. The TOTAL

annunciator will light during the TOTAL weight display. After five seconds of Total

Weight display, the number of samples is displayed for two seconds.

Advertising