If you are already in the Cal menu
skip to step 2

1)

Remove the seal screw. With
a small screwdriver press the
Cal Switch behind the seal
screw. The CAL Setup Menu
appears.

2)

Press the

USER

key to

scro ll to ”Setup” .

3)

Press

TARE

to enter the Cal

Setup Menu.

4)

Press

USER

to scroll to the

“Auto0” menu.

5)

Press

TARE

to enter the Auto-

Zero Menu. The current setting
is displayed (blinking)

6)

Change the setting between
On or OFF using the

USER

key.

7)

Press

TARE

to set the

Autozero. The display goes to
the next item in the CAL Setup
Menu.

8)

Either press

ZERO

twice to

exit Setup and store all
changes, or continue to
another Setup Menu item
using the

USER

Key.

&$/

Auto Zero
Maintenance Setup

VHWYS

6WDQG

$YWR

2II

6WDQG

Next Cal Setup Menu Item

Store and return to

weight display

&D/

6WRUH

Calibrate Menu

Blinking

2Q

Blinking

USER

F

Remove
Hex Seal
Screw

Cal Switch

Access

Press Switch
behind the hole
with a small
Screwdriver

Calibration 41

Figure 5-10. Auto Zero Maintenance