Xq/g, C-cal calibration procedure – Rice Lake MSI-4260 Port-A-Weigh User Manual

Page 41

Advertising
background image

XQ/G



&&$/



1) Remove the seal screw. With

a small screwdriver press the
Cal Switch behind the seal
screw.

The CAL Setup Menu
appears.

2) Press the

USER

key to

scroll to the C-Cal menu
selection.

3) Press

TARE

to Start the C-

Cal Procedure.

If this calibration is using a cleared

4260, the INITIAL CALIBRATION

procedure will start here. You must

set the cal unit, capacity, and count-

before proceding to step 4,

4) The display reads “UnLd”

(unload) indicating you should
remove all weight from the
hook.

5) Press

TARE

. The 4260 sets

the zero calibration point.

6) if the zero is in range, the scale

will display “PASS”.

7) Next the display reads

“CCAL?” to ask that you look
up the C-Cal value to be used.

8) Press

TARE

. The 4260 is

ready for numeric entry of the
C-CAL value.

9) To enter an C-Cal value

press

USER

. The display

far left digit will flash zero
indicating that a number
should be entered.

&$/



C-Cal Calibration
Procedure

blinking

3$66

No weight

on hook

Flashes

Zero

Input C-Cal value

blinking

&&D/

TARE

T

TARE

T

TARE

T

USER

F

USER

F

Remove
Hex Seal
Screw

Cal Switch

Access

Press Switch
behind the hole
with a small
Screwdriver

blinking

Calibration 37

Figure 5-6. C-Cal Calibration Procedure