TARE

T

TARE

T

TARE

T

ZERO

O

sto

ZERO

O

sto

USER

F

USER

F

If you are already in the Cal menu
skip to step 2

1)

Remove the seal screw (cut
lead wire seal if present). With
a small screwdriver press the
Cal Switch behind the seal
screw.

The CAL Setup Menu
appears.

2)

Press the

USER

key to

scroll to ”Setup” .

3)

Press

TARE

to enter the Cal

Setup Menu.

4)

Press

TARE

to enter the

Standard Menu. The current
standard setting is displayed.

5)

Scroll to the desired standard
using the

USER

key. .

6)

Press

TARE

to set the

standard. The display goes to
the next item in the CAL Setup
Menu.

7)

Either press

ZERO

twice to

exit Setup and store all
changes, or continue to
another Setup Menu item
using the

USER

Key.

&$/

LFT Standard Setup

VHWYS

6WDQG

LQGXV

XQLW

$YWR

Next Cal Setup Menu Item

Store and return to

weight display

&D/

6WRUH

Calibrate Menu

Remove
Hex Seal
Screw

Cal Switch

Access

Press Switch
behind the hole
with a small
Screwdriver

40 MSI-4260 Port-A-Weigh Operator’s Manual

Figure 5-9. LFT Standard Setup

5.5 Auto Zero Maintenance

The MSI-4260 employs an autozeroing maintenance mechanism to adjust the zero

reading to the Center-of-Zero (COZ) COZ is defined as the weight reading is within

1/4 ‘d’ of zero. AZM continuously adjusts zero to maintain COZ. It is recommended

that AZM is on to maintain the highest accuracy. However, there are circumstances

when it should be turned off. This can happen when minor variations of weight occur

while picking up scale attachments and the variations fall within the AZM capture

window. The AZM capture window (usually 1 ‘d’) and capture time (usually 8

seconds) can be adjusted by MSI to meet custom requirements. The settings of AZM

are dictated in Legal-for-Trade Standards and cannot be adjusted.