1 standard, Standard – Rice Lake MSI-4260 Port-A-Weigh User Manual

Page 43

Advertising
background image

&$/

&&D/

6HWYS

6WDQG

$YWR

LQGYV

+E

U

XQLW

2Q

2II

Cal Standard

Auto Zero

Maintenance

Calibrate with Weights

Calibrate with Constant

Industrial Setting

(default)

Handbook 44, NTEP

R-76, OIML

Industrial, restricted to
1 Weight Unit

AZM On

AZM Off

Remove
Hex Seal
Screw

Cal Switch

Access

Press Switch
behind the hole
with a small
Screwdriver

USER

F

Calibration 39

Figure 5-8. Calibration Setup Menu Map

5.4.1

Standard

There are 4 selections in the standards menu.

1. Industrial (indu5) – This is the most common setting for the 4260. With the

Industrial standard, you have full range zero, access to units switching, filters,

and peak hold.

2. Handbook 44 (Hb-44) – Sets the scale to enable only approved features per the

NTEP HB-44 rules and regulations. Access is denied to Peak Hold, and the zero

range may be limited. The Filter menu is moved to the Cal Setup Menu, so

filters are only accessible through the Cal Seal.

3. R-76 (r-76) – Sets the scale to enable only approved features per OIML R-76.

Only kg weight units are available. The zero range is limited to 5% (-2 +3%

relative to Calibrate zero). Net/Gross function is temporary. once Net weight is

established, pushing an F key set for Net/Gross will cause a maximum 5 second

display of the Gross weight. You must clear the Tare to display Gross weight

constantly. Other metrological aspects are changed to meet R-76 requirements.

4. One Unit (1unit) – The one unit Standard is exactly the same as Industrial,

except units switching is inhibited. This is useful for Metric only countries.

Another use of the One Unit standard is to allow the scale to be calibrated in

units other than lb or kg, since conversions are eliminated.

Contact MSI for more information on the Standards settings.