15) The display shows “CAL’d”

to indicate that the
calibration was successful.

16) Press

TARE.

The display

flashes “C-CAL” followed
by the C-CAL number.

Make a note of this number.

If the X1000 annunciator is lit,

also make a note of that.
This number is used to calibrate when

test weights are unavailable. C-CAL

does not replicate multipoint

calibrations.

17) Press

ZERO

to store the

calibration constants, or
after 8 seconds,the
calibration constants will
automatically stored. After
the Store message is
displayed, the scale goes
to the next item on the Cal
Menu, “SEtuP”.

18) Press

ZERO

to exit the Cal

menus and start up the
standard weight display.
The 4260 is ready for use.

19) After ensuring the

calibration is acceptable,
replace the Hex Seal Screw.

&&D/



Example Value only.

VWRUH

&$/G

6HWYS

Flashing

ZERO

O

sto

ZERO

O

sto

ZERO

O

sto

TARE

T

ZERO

O

sto

If the scale is being used as a Legal-

For-Trade device, place a Lead-

Wire Seal through the Cal Side

Bored Hex Head Screw over to the

adjacent Side Bored Cap Screw.

To skip the

C-CAL step

14) The display now reads

“LoAd3” or “LoAd4”.
Repeat steps 12 and 13
for additional span points
press

TARE

or press

ZERO

if you are finished.

After LoAD4 has passed, the 4260

will automatically advance to the next

step.

/RDG

/RDG

or

To enter another

span point

TARE

T

Calibration 33

Figure 5-3. Standard Calibration Procedure (continued)