54 MSI-4260 Port-A-Weigh Operator’s Manual

Display toggle between
“Error” and “UnLd”

Weight in below the zero
range.

If scale is in compression, remove
the source.

Calibration faulty

Recalibrate

Faulty Load Cell or wiring Check load cell connections

Display toggles
between “Error” and
“A2DLo”

A/D is saturated negative

Check load cell and load cell
wiring

Display toggles
between “Error” and
“buttn”

A key is stuck or is being
held down

Check switches for damage
Ensure that a remote is not
transmitting continuously

RF Remote does not
work

Units not mated

See Section 6.3 and Section 6.4.

Some RF Remote keys
do not work but the
ACK light blinks

Keys were not enabled
during the setup process

Enable keys by running the
transmitter and receiver address
procedures

Lo Batt is blinking

Battery is low

Recharge Battery

Unit turns on, then
immediately off

Battery is low

Recharge Battery

Weight will not zero

System not stable

Stable annunciator must turn on
for Zero to function. Increase
filtering for more stability.
Increase filtering for more stability

Zero out of range

LFT units have limited zero range.
Reduce the weight or use Tare
instead

Weight will not Zero,
Tare or Total

System is not stable

Wait for Stable annunciator to
turn on, or if in a mechanically
noisy crane, increase the filtering
or increase the size of the scale
increment ‘‘d’. It is also possible
to increase the motion window.
Contact MSI if you have a
problem getting the MSI-4260 to
zero, tare, or total due to stability
issues.

Set Point Lights Blink

Set point is enabled and
the trigger point has been
reached

Disable set points if they are not
needed

Manual Total does not
work

A Function key is not set
to “Total”

Set up Func1 or Func2 for “Total”

Weight must be stable

Increase filtering for more stability

Problem

Possible Cause

Solution

Table 8-1. Troubleshooting