6 instrument commissioning, 1 if the instrument has not been calibrated – Rice Lake SCT-10 User Manual

Page 12

Advertising
background image

8

SCT Weight Transmitter Operator’s Manual

1.6

Instrument Commissioning

1. Plug power cord in to outlet to turn on indicator, the display shows in

sequence:
- “SU” followed by the software code (e.g.: SU S );
- - “r” followed by the software version (e.g.: r 1.04.01 );
- - “HU” followed by the hardware code (e.g.: HU 104 );
- - the serial number (e.g.:1005 15 );

2. Check that the display shows the weight and that when loading the load

cells there is an increase in weight.

3. If there is not, check and verify the connections and correct positioning

of the load cells.

If instrument has NOT been calibrated complete Section 2.1 before
proceeding to next step.

4. Reset to zero. See Section 2.1.3 “Zero Setting” on page 15.
5. Check the calibration with test weights and correct the indicated weight

if necessary. See Section 2.1.5 “Weight (Span) Calibration (With Test

Weights)” on page 16.

6. If you use the analog output, set the desired analog output type and the

full scale value. See Section 2.4 “Analog Output” on page 21.

7. If you use serial communication, set the related parameters. See Section

2.5 “Serial Communication Settings” on page 23.

8. If setpoints are used, set the required weight values and the relevant

parameters. See Section 2.8 “Setpoints Programming” on page 29 and

Section 2.6 “Outputs And Inputs Configuration” on page 26.

1.6.1 If The Instrument Has Not Been Calibrated

Missing plant system identification tag, proceed with calibration:

1. If load cells data are unknown, follow the procedure in Section 2.1.5

“Weight (Span) Calibration (With Test Weights)” on page 16.

2. Enter the rated data of load cells following the procedure given in

Section 2.1.1 “Theoretical Calibration” on page 14.

Note