1 calibration, Ted complete section 2.1 – Rice Lake SCT-10 User Manual
Page 15
Configuration 11
2.1
Calibration
Figure 2-2. Calibration Menu Structure
Parameter
Choices
Description
FS-TED
Enter #
deno *
System Full Scale is determined by multiplying one
load cell capacity by the number of load cells used.
Example of system full scale value calculation:
4 cells of 1000kg ---->
FULL SCALE = 1000 X 4 = 4000
The instrument is supplied with a
theoretical full scale value deno
corresponding to 10000. To restore
factory values, set 0 as full scale.
SENS LB
Enter #
0.50000
to
7.00000
2.00000 *
Sensitivity is a load cell rated parameter expressed
in mV/V. Set the average sensitivity value indicated on
the load cells. I
Example of 4-cell system with sensitivity
2.00100, 2.00150, 2.00200, 2.00250;
enter 2.00175,
calculated as (2.00100 + 2.00150 + 2.00200 +
2.00250) / 4.