Configuration 11

2.1

Calibration

Figure 2-2. Calibration Menu Structure

Parameter

Choices

Description

FS-TED

Enter #
deno *

System Full Scale is determined by multiplying one
load cell capacity by the number of load cells used.
Example of system full scale value calculation:

4 cells of 1000kg ---->

FULL SCALE = 1000 X 4 = 4000

The instrument is supplied with a
theoretical full scale value deno
corresponding to 10000. To restore
factory values, set 0 as full scale.

SENS LB

Enter #
0.50000
to
7.00000

2.00000 *

Sensitivity is a load cell rated parameter expressed
in mV/V. Set the average sensitivity value indicated on
the load cells. I
Example of 4-cell system with sensitivity

2.00100, 2.00150, 2.00200, 2.00250;

enter 2.00175,

calculated as (2.00100 + 2.00150 + 2.00200 +

2.00250) / 4.

Table 2-2. Calibration Menu