38 SCT Weight Transmitter Operator’s Manual

Weight (Span) Calibration Commands (With Test Weights)

The instrument calibration can be changed via MODBUS. To carry out this

procedure, the system must be unloaded and the weight value display reset to

zero with the command ‘100’ of the Command Register. Then, a load must be

placed on the system and the correct weight value must be sent to the registers

40037-40038; to save this value, send the control ‘101’ from the Command

Register. If the operation is successfully completed, the two test weight registers

are set to zero.

Analog Output Setting

Write the weight in the registers “Weight value corresponding to the Full Scale of

analog output H” (40045) and “Weight value corresponding to the Full Scale of

analog output L” (40046) or write the weight in the registers “weight value

corresponding to the ZERO of the analog output H” (40043) and “weight value

corresponding to the ZERO of the analog output L” (40044). After writing the

value, send the command 99 from the Command Register to save it to EEPROM
memory.

Status Register

Bit 0

Cell Error

Bit 1

AD Convertor Malfunction

Bit 2

Maximum weight exceeded by 9 divisions

Bit 3

Gross weight higher than 110% of full scale

Bit 4

Gross weight beyond 999999 or less than -999999

Bit 5

Net weight beyond 999999 or less than -999999

Bit 6

Bit 7

Gross weight negative sign

Bit 8

Net weight negative sign

Bit 9

Peak weight negative sign

Bit 10

Net display mode

Bit 11

Weight stability

Bit 12

Weight within +/-¼ of a division around ZERO

Bit 13

Bit 14

Bit 15