Configuration 29

2.8

Setpoints Programming

1. Press

to enter setpoints and hysteresis settings.

2. Press

or

until desired setpoint or hysteresis parameter is

displayed, press

.

3. Press

or

until desired value is displayed, press

.

4. Press

to exit setpoints and hysteresis settings.

These values are set to zero if the calibration is changed
significantly (see Section 2.1.1 “Theoretical Calibration” on
page 14
and Section 2.1.5 “Weight (Span) Calibration (With Test
Weights)” on page 16).

Parameter Choices

Description

In

N/A

Input Test - for each open input 0 is displayed, 1 is
displayed when the input is closed.

Out

0 *
1

Output Test -
Setting 0 - the corresponding output opens.
Setting 1 - the corresponding output closes.

ANALOG ANALOG

Allows the analog signal to range between the minimum
and the maximum values starting from the minimum.

NA

current output test

UOLt1

voltage output test

NU-CEL

N/A

Millivolt Test - displays the load cell response signal in
mV with four decimals.

*

- indicates default value.

Table 2-9. Test Menu

Parameter

Choices

Description

SEtP 1
SEtP 2

0-Full Scale
0 *

Setpoint; relay switching occurs when the weight
exceed the value set in this parameter. The type of
switching is settable (see “Outputs And Inputs
Configuration” on page 26).

HYStE 1
HYStE 2

0-Full Scale
0 *

Hysteresis, value to be subtracted from the setpoint to
obtain contact switching for decreasing weight. For
example with a setpoint at 100 and hysteresis at 10,
the switching occurs at 90 for decreasing weight.

*

- indicates default value.

Table 2-10. Setpoints

MENU

TARE

PRINT

MENU

TARE

PRINT

MENU

0

ESC

Note