2 filter on the weight – Rice Lake SCT-10 User Manual

Page 23

Advertising
background image

Configuration 19

2.2

Filter On The Weight

The filtering selection is used to eliminate environment noise, and is typically a

compromise between responsiveness and stability. The lower the number, the

more responsive the display will be to weight changes. The filter is used to

stabilize a weight as long as the variations are smaller than the corresponding

“Response Time”. The filter setting is dependent on the type of application and

the required update rate.
Setting this parameter allows a stable weight display to be obtained. To increase

the effect (weight more stable), increase the value.

1. Press and hold

, then press

. CALIb will be displayed.

2. Press

or

until FILtEr is displayed, press

. LED’s will

begin scrolling.

3. Press

or

until desired filter value is displayed, press

.

4. The weight is displayed (all LED’s will be flashing) and the displayed

stability can be experimentally verified. Press

.

5. If stability is not satisfactory, press

, this returns indicator to FILtEr

option and the filter may be modified again until an optimum result is

achieved.

6. Press

to exit set-up menu.

The filter enables to stabilize a weight as long as its variations are
smaller than the corresponding “Response Time”. It is necessary
to set this filter according to the type of application and to the full
scale value set.

Filter Value

Response times

[ms]

Display and serial port refresh frequency

[Hz]

0

80

80

1

190

80

2

260

40

3

450

26

4*

900

13

5

1700

13

6

2500

13

7

4200

10

8

6000

10

9

7500

5

* indicates default value.

Table 2-4. Filter Settings

MENU

0

ESC

TARE

PRINT

MENU

TARE

PRINT

MENU

MENU

MENU

0

ESC

Note

Advertising