SCT Weight Transmitter Operator’s Manual

1.2.2 Load Cell Testing

Load Cell Resistance Measurement (Use A Digital Multimeter):

• Disconnect the load cells from the instrument and check that there is no

moisture in the load cell junction box caused by condensation or water

infiltration. If so, drain the system or replace it if necessary.

• The value between the positive signal wire and the negative signal wire must

be equal or similar to the one indicated in the load cell data sheet (output

resistance).

• The value between the positive excitation wire and the negative excitation

wire must be equal or similar to the one indicated in the load cell data sheet

(input resistance).

• The insulation value between the shield and any other load cell wire and

between any other load cell wire and the body of the load cell must be higher

than 20 Mohm (mega ohms).

Load Cell Voltage Measurement (Use A Digital Multimeter):

• Remove weight of scale from load cell to be tested.
• Make sure that the excitation wires of the load cell connected to the

instrument is 5 Vdc +/- 3%.

• Measure the millivolt signal between the positive and the negative signal

wires by directly connecting them to the multi-meter, and make sure it reads

between 0 and 0.5 mV (thousandths of a Volt).

• Apply load to the load cell and make sure that there is a signal increment.

If one of the above conditions is not met, please contact the
technical assistance service.

1.3

Specifications

• Weight indicator and transmitter for Omega/DIN rail mounting suitable for

back panel; space saving vertical shape. Six-digit semi alphanumeric display

(18mm h), 7 segment. Four-key keyboard. Dimensions: 25x115x120 mm.

• Displays the gross weight; with an external contact capable of remote

zeroing and gross/net switching.

• IP67 box version (dimensions: 170x140x95mm). Four fixing holes diameter

4mm (center distance 122x152mm).

• Peak weight function.

Important

ANALOG OUTPUT

(Current and Voltage)

PLC

PLC or FIELD SIGNALS

2 DIGITAL INPUTS

(Optoisolated,
Externally supplied)

2 RELAY

OUTPUTS

MAX 8 LOAD CELLS IN

PARALLEL

RS-485
RS-232
Modbus RTU

SERIAL PORT

DC power

supplier

(12-24 Volt )