30 SCT Weight Transmitter Operator’s Manual

2.9

Reserved For The Installer

2.9.1 Default Scale

Operation must only be performed after contacting technical
assistance

1.With power off, press and hold

, then power on.

Display shows PrOG .

2. Press

, display shows bASE.

3. Press

, display shows UAIt.

4. Instrument will reboot.

By confirming the displayed program, the system variables are set
with default values.

2.9.2 Program Selection - Reverse:

Scale capacity is displayed when scale is empty. As weight is added display will

count down.

1. With power off, press and hold

, then power on.

Display shows PrOG .

2. Press

, display shows bASE.

3. Press

, display shows rEuEr.

4. Press

, display shows UAIt.

5. Instrument will reboot.

By pressing

you will quit the program without introducing any changes and

without deleting any of the set variables.

If you do not have a specific manual for the newly set program, you
can request it from technical assistance.