3 zero parameters – Rice Lake SCT-10 User Manual
Page 24
20 SCT Weight Transmitter Operator’s Manual
2.3
Zero Parameters
1. Press and hold
, then press
. CALIb will be displayed.
2. Press
or
until PArA 0 is displayed, press
.
3. Press
or
until desired parameter (see Table 5 ) is displayed,
press
. The currently programmed value is displayed and LED’s
will be scrolling.
4. Press
or
until desired value is displayed, press
.
5. Press
twice to exit set-up menu.
Parameter
Choices
Description
0 SEt
Enter #
0-max full scale
300 *
Considered decimals: 300
– 30.0 – 3.00 – 0.300
Resettable Weight setting for small
weight change.
Indicates the maximum weight value
resettable by external contact, keypad or
serial protocol
AutO 0
Enter #
0 - max 20% of full scale
0 *
Automatic zero setting at power-on
If when indicator is powered on the
weight value is lower than the value set
in this parameter and does not exceed
the 0 SEt value, the weight is reset.
To disable this function set to 0.
trAC 0
nOnE *
1-5
Zero tracking
When the zero weight value is stable
and, after a second, it deviates from zero
by a figure in divisions smaller or equal to
the figure in divisions set in this
parameter, the weight is set to zero.
To disable this function
,
set to none
Example: if the parameter dI UI
S
is
set to 5 and trAC 0 is set to 2, the
weight will be automatically set to
zero for variations smaller than or
equal to 10 (dI UI S x trAC 0 ).
*
- indicates default value.
Table 2-5. Zero Parameters Settings