3 zero parameters – Rice Lake SCT-10 User Manual

Page 24

Advertising
background image

20 SCT Weight Transmitter Operator’s Manual

2.3

Zero Parameters

1. Press and hold

, then press

. CALIb will be displayed.

2. Press

or

until PArA 0 is displayed, press

.

3. Press

or

until desired parameter (see Table 5 ) is displayed,

press

. The currently programmed value is displayed and LED’s

will be scrolling.

4. Press

or

until desired value is displayed, press

.

5. Press

twice to exit set-up menu.

Parameter

Choices

Description

0 SEt

Enter #
0-max full scale
300 *
Considered decimals: 300
– 30.0 – 3.00 – 0.300

Resettable Weight setting for small
weight change.
Indicates the maximum weight value
resettable by external contact, keypad or
serial protocol

AutO 0

Enter #
0 - max 20% of full scale
0 *

Automatic zero setting at power-on
If when indicator is powered on the
weight value is lower than the value set
in this parameter and does not exceed
the 0 SEt value, the weight is reset.
To disable this function set to 0.

trAC 0

nOnE *
1-5

Zero tracking
When the zero weight value is stable
and, after a second, it deviates from zero
by a figure in divisions smaller or equal to
the figure in divisions set in this
parameter, the weight is set to zero.
To disable this function

,

set to none

Example: if the parameter dI UI

S

is

set to 5 and trAC 0 is set to 2, the
weight will be automatically set to
zero for variations smaller than or
equal to 10 (
dI UI S x trAC 0 ).

*

- indicates default value.

Table 2-5. Zero Parameters Settings