Introduction

3

Windy Conditions - Shocks - Vibrations:

The use of weigh modules is strongly recommended for all load cells to

compensate for misalignment of the support surfaces. The system designer must

ensure that the scale is protected against lateral shifting and tipping relating to

shocks and vibration, windy conditions, seismic conditions and stability of the

support structure.

Grounding The Weighed Structure:

By means of a 10ga solid or braided wire or braided grounding strap, connect the

load cell upper support plate with the lower support plate, then connect all the

lower plates to a single earth ground. Once installed electrostatic charges

accumulated are discharged to the ground without going through or damaging the

load cells. Failure to implement a proper grounding system might not affect the

operation of the weighing system; this, however, does not rule out the possibility

that the load cells and connected instrument may become damaged by ESD. It is

forbidden to ensure grounding system continuity by using metal parts contained

in the weighed structure.(see Figure 1-1.)

Figure 1-1. Installation Recommendations

1.2

Load Cells

1.2.1 Load Cell Input Test (Quick Access)

1. From the weight display, press

for 3 seconds.

2. The display will read NU-CEL. Press

.

3. The response signal of the load cell is displayed, expressed in mV with

four decimals. Press

three times to exit set-up mode.

Uses ground plate

to continue ground.

Uses structure to continue ground.

PRINT

MENU

0

ESC