5 serial communication settings, Ters. see section – Rice Lake SCT-10 User Manual
Page 27
Configuration 23
2.5
Serial Communication Settings
Figure 2-3. Serial Communications Menu Structure
According to the chosen protocol only the necessary settings will be displayed
in sequence.
Parameter
Choices
Description
rS-485
(Communitcation
Port)
None *
Disables any type of communication (default).
NOdbuS
MODBUS-RTU protocol; possible addresses: from
1 to 99 (see Section 3.6)
ASCII
ASCII bidirectional protocol; possible addresses:
from 1 to 99 (see Section 3.7)
NOdU6-
NOd td
COntIn
Continuous weight transmission protocol (see
Section 3.8), at the frequency set in HERTZ
parameter (from 10 to 300).
NOd t(set: PARITY=none, STOP=1)
NOd td(set: PARITY=none, STOP=1)
rI P
Continuous weight transmission protocol, streams
net and gross (see Section 3.9)
(set: BAUD=9600,PARITY=none, STOP=1)
Hdr1 P
Continuous weight transmission protocol, streams
net and gross including decimal (see Section 3.9)
(set: BAUD=9600,PARITY=none, STOP=1)
Table 2-7. Serial Communications Menu