5 serial communication settings, Ters. see section – Rice Lake SCT-10 User Manual

Page 27

Advertising
background image

Configuration 23

2.5

Serial Communication Settings

Figure 2-3. Serial Communications Menu Structure

According to the chosen protocol only the necessary settings will be displayed
in sequence.

Parameter

Choices

Description

rS-485
(Communitcation
Port)

None *

Disables any type of communication (default).

NOdbuS

MODBUS-RTU protocol; possible addresses: from
1 to 99 (see Section 3.6)

ASCII

ASCII bidirectional protocol; possible addresses:
from 1 to 99 (see Section 3.7)

NOdU6-
NOd td

COntIn

Continuous weight transmission protocol (see
Section 3.8), at the frequency set in HERTZ
parameter (from 10 to 300).
NOd t(set: PARITY=none, STOP=1)
NOd td(set: PARITY=none, STOP=1)

rI P

Continuous weight transmission protocol, streams
net and gross (see Section 3.9)
(set: BAUD=9600,PARITY=none, STOP=1)

Hdr1 P

Continuous weight transmission protocol, streams
net and gross including decimal (see Section 3.9)
(set: BAUD=9600,PARITY=none, STOP=1)

Table 2-7. Serial Communications Menu

Addr

bAud

HErt2

dELAY

PArItY

StOP

rS485

nOnE

NodbUS

ASCII

Cont In

rI P

HdrI P

HdrI Pn

Enter #

9600

19200

38400

115200

4800

2400

30
40

50

60

20

10

80

70

Enter #

EUEn

Odd

nOnE

2
3

4

5

1

0

8
9

7

6

CALIb

FILtEr

PArA 0

SErIAL

tESt

Out-In

ANA LOG