Operation 43
Error messages:
In case of an instrument alarm for exceeding 110% of the full scale or 9 divisions
above the value of the parameter NASS, the instrument sends the string:
&aassO-Lst\ckck
In case of faulty connection of the load cells or of another alarm, the instrument
sends:
&aassO-Fst\ckck
in which:
s = 1 separator character (32 ASCII – space-).
Refer to Section 3.5 “Alarms” on page 33.
Semi-Automatic Zero (Weight Zero-Setting For Small Variations)
The zero-setting will not be maintained after an instrument
power-off.
The PC transmits: $aaZEROckckCR
Possible instrument responses:
- correct reception: &&aa!\ckckCR
- incorrect reception: &&aa?\ckckCR
- the current weight is over the maximum value resettable: &aa#CR
Switching From Gross Weight To Net Weight
The PC transmits: $aaNETckckCR
Possible instrument responses:
- correct reception: &&aa!\ckckCR
- incorrect reception: &&aa?\ckckCR
Switching From Net Weight To Gross Weight
The PC transmits: $aaGROSSckckCR
Possible instrument responses:
- correct reception: &&aa!\ckckCR
- incorrect reception: &&aa?\ckckCR
Reading Of Decimals And Number Of Divisions
The PC transmits: $aaDckckCR
Possible instrument responses:
- correct reception: &aaxy\ckckCR
- incorrect reception: &&aa?\ckckCR
in which:
x = number of decimals
y = division value
The y field acquires the following values:
'3' for division value = 1
'4' for division value = 2
'5' for division value = 5
'6' for division value = 10
'7' for division value = 20
Important