12 SCT Weight Transmitter Operator’s Manual
dI UI S
1
2 *
5
10
20
50
100
0.0001
0.0002
0.0005
0.001
0.002
0.005
0.01
0.02
0.05
0.1
0.2
0.5
Division (resolution) - the weight increment (display
division size) that the scale counts by.
Selections are: 0.0001 and 100 with x1 x2 x5 x10
increments.
NASS
Enter #
0 *
to
max full scale
Maximum capacity (Live Load/Product) that can
be displayed. When the weight exceeds this value by
9 divisions, the display will go to dashes, indicating
overload.
Setting this value to 0 will disable the over capacity
function.
ZErO
0
Used to capture the deadload of the scale
system. With the scale empty, the displayed value
can be zeroed off. This menu may also be accessed
directly from the weighing mode to compensate for
zero changes or variations.
Press
to display the accumulated weight that
has been zeroed off.
INP 0
Enter #
0 to 999999
0 *
Estimated dead load value of the scale when a
scale contains product that cannot be removed. The
value entered is the dead load. This value will be
replaced if the zero function is performed later.
Parameter
Choices
Description
Table 2-2. Calibration Menu
TARE