3 zero setting, 4 zero value manual entry, 3 zero setting 2.1.4 zero value manual entry – Rice Lake SCT-10 User Manual

Page 19

Advertising
background image

Configuration 15

2. Press ,

FS-tEO is displayed.

3. Press

or

until

NASS is displayed, press

. LED’s will

begin scrolling.

4. Press

or

until desired capacity is displayed, press

.

5. Press

twice to exit set-up menu.

2.1.3 Zero Setting

Perform this procedure after having set the “Theoretical Calibration” on page 14.

This menu may also be accessed directly from the weight display,

press and hold

for 3 seconds.

1. Press and hold

, then press

. CALIb will be displayed.

2. Press ,

FS-tEO is displayed.

3. Press

or

until ZErO is displayed, press

.

4. The weight value to be set to zero is displayed. In this phase all of the

LEDs are flashing. Press

, the weight is set to zero (the value is

stored to the permanent memory).

5. Press

twice to exit set-up menu.

Press

to display the accumulated deadload that has been

zeroed off by the instrument, displaying the sum of all of the
previous zero settings.

2.1.4 Zero Value Manual Entry

Perform this procedure only if it is not possible to reset the
weighed structure tare, for example because it contains
product that can not be unloaded.

Set in this parameter the estimated zero value.

1. Press and hold

, then press

. CALIb will be displayed.

2. Press ,

FS-tEO is displayed.

3. Press

or

until

INP 0 is displayed, press

. LED’s will

begin scrolling.

4. Press

or

until desired dead load is displayed, press

.

5. Press

twice to exit set-up menu.