3. Once the pan has been installed, level the scale by adjusting the rear feet until the level indicator

shows the “air bubble” is in the center position of the sight glass. The level indicator is located in the
rear left side of the scale.


4. After leveling the scale, you may plug the AC adaptor to the AC adaptor socket located in the rear of

the scale.

5. When the AC adapter is plugged into the wall outlet, the scale will turn “ON” automatically, initialize

itself, perform automatic internal calibration and enter weighing mode.


6. To put the scale into standby mode, leave the AC adaptor plugged into both the scale and the wall

outlet and press the Power “OFF” button (

). The “OFF” indicator will light up in the upper left

corner of the display signaling the scale is in standby mode.