- Digital Fiber Sensor FS-N10 Series User's Manual -

Zero shift calibration

This sensitivity setting performs the zero shift function and basic calibration (two-point
calibration / maximum sensitivity calibration / full auto calibration) simultaneously.
The lower of the received light intensity values specified at the time of sensitivity
setting will automatically be set to "0".

The light intensity received with a workpiece present will be applied with
the same amount of correction as for that received with no workpiece.

The following basic calibrations can be set while in the zero shift calibration.

• Two-point calibration
• Maximum sensitivity calibration
• Full auto calibration

If two-point calibration is performed with a reflective model when the light
intensity received with a workpiece is "2500" and without a workpiece is
"500":

Reference

Setting example

After two point calibration

(Before zero shift calibration)

Zero shift calibration is executed

500

2500

500

2500

0

2000

1000

Setting
value

Setting
value

1500

The light intensity received
with no workpiece is
subtracted from the whole.

The state with no
workpiece (the lowest of
received light intensity) is
set to "0".

Received light intensity

Received light intensity