Removing and replacing the, Stack temperature limit switch – LAARS NeoTherm NTV (Sizes 399-850 MBTU/h) - Service Manual User Manual

Page 57

Advertising
background image

47

Service Manual - NeoTherm 500

The stack temperature limit switch will shut down the

NT unit if the stack temperature rises above 195°F.

This is a “duplex” sensor - it actually includes two

separate thermistors. The control system constantly

compares the readings from the thermistors. If there

is a large difference between the readings, or if one

of the thermistors is “open” or “shorted,” the control

system will shut down the NT unit, and display a

“Hold 95” error message.
Tools and equipment required:
• Crescent wrench or 17 mm socket
Procedure:
1. Turn off power to the system. Use the main dis-

connect switch mounted above the front panel.

2. The stack temperature limit switch is mounted in

the “far” end of the heat exchanger assembly (the

end which is farthest from the operator control

panel.) You should be able to reach this switch

from the top of the unit. Remove the top panels

so you can see the switch.

3. Unplug the wiring connector (four wires). The

connector is locked in place by a “squeeze tab.”

Press on the tab to release the connector.

4. Unscrew the sensor using the 17 mm socket or

a crescent wrench. Turn counter-clockwise to

unscrew the part.

5. To reassemble, reverse the process we have just

described. There is no need to use pipe dope or

Teflon® tape on the threaded part of the sensor.

Fig. B25-1 - Stack temperature limit switch (typical)

B25

Removing and Replacing the Stack Temperature Limit Switch

Advertising
This manual is related to the following products: