User auto test, Summary – EXFO MaxTester 635 User Manual

Page 236

Advertising
background image

Copper Auto Tests

228

MAX-635

User Auto Test

User Auto Test

User Auto Test runs a range of different user-defined tests that compare
measured results against stored threshold values to provide pass or fail
status and results. Test parameters and thresholds are inherited from
individual tests.

To access User Auto Test:
1.
From the Copper Main menu, use the up/down, left/right arrow keys

to highlight Auto Test and press

.

2. Highlight User Auto Test and press

to start the test.

Summary

The Summary page displays the overall pass/fail
status of all tests selected for auto testing.

Current Group is the current test group name, for
example Auto Test.

Current Test displays the test being measured.
Here you can switch to another test from the given
list. However, this action will terminate the
ongoing measurement test.

When FED is Enabled, the FED icon and numeric
state are displayed beside the Current Test.

Test Selection button opens a new page that allows you to select the
tests to include in the User Auto Test.

Result displays each test selected and its status.

Test Status displays the current state of measurement or an error
message.