Page 19

Advertising
background image

2.

PRELIMINARY OPERATIONS

2.1 Select Instrument Configuration

(Language, Microphone Response Etc.)

Select the language in which the instrument will present menu options.
This should be the first operation so that the displays and options will be

understood.

Similarly, to get the most accurate results, these instruments

must be operated with a known microphone directional response and be
calibrated for this response.

When the language and microphone response are known to be

correct for the proposed measurement task, go directly to Section 2.2

Calibration Check.

European IEC standards require measurements to be performed

with microphones that have a Free Field response, while the U.S.A.
ANSI standards require Random Incidence microphones. The micro-

phone response is set via the Configuration menu.

Normally, these more permanent characteristics must be set

only once, as the instrument will store the settings and use them the
next time it is switched on.

Proceed as follows.

Operation Press

Display Shows

Comments

1. Switch the

instrument

ON

Start up messages are

displayed that include:

Instrument Type,

Firmware Version, Etc.

Finally, the Configuration

screen is displayed with

Battery Voltage, followed by

the Calibration check screen.

2. Select the

Main menu

2 times

This is the Main menu.

It gives access to further

menus that select:

Measurement Mode,

Measurement Setup,

Calibration Sequence,

Timers

Data Recall from Memory,

Instrument Configuration,

Status.

CEL-450/490 Operators Handbook - Page 19

Preliminary Operations