Select mode & setup – Casella CEL CEL-450 User Manual

Page 35

Advertising
background image

Operation Press

Display Shows

Comments

9. Select

Period

from the

Setup

screen to

enable

measure-

ment

intervals to

be set

or

then

This screen allows a period

to be specified for regular

sequential measurement.

Period measurement is

available only with a

CEL-490.

The predicted maximum run

time with the currently

selected parameters and

periods and the available

unused memory are also

indicated.
The narrowband measure-

ment period can be set to:

10, 20, 50, 100, 250, 500

milliseconds,

1, 5, 10, 15, 20, 30 seconds,

1, 5, 10 ,15, 20 , 30 minutes,

1 hour, - - - - (=none).

Press

to access quick

edit to specify measurement

and periods.

10. Select

Period

Functions

from the

Period

Setup

screen

This screen allows the

following measurement

functions to be specified

(

=enabled, X =disabled)

using the weightings set in

step 5 above:

L

AF

L

AF50.0

.

L

AFmx

L

AF90.0

L

AFmn

L

AF95.0

L

Aeq

L

AF99.0

L

AF10.0

.

Note that L

Zpk

is not

available as a period

function.

11. If required,

enable (

)

or disable

(X) the

functions

to

enable

quick

edit

mode

Use

and

to select a

field.
Use

and

to change

the entry in the field.

Save the settings by

pressing

again to

resume normal operation.

CEL-450/490 Operators Handbook - Page 35

Select Mode & Setup

Advertising