Acoustic calibration check – Casella CEL CEL-450 User Manual

Page 40

Advertising
background image

Operation Press

Display Shows

Comments

5. When the

calibration

is

successful

The user is given the option

of saving or not saving the

new calibration.

Press

again to use the

highlighted option and save

the calibration.

Or use

or

to

highlight NO SAVE to ignore

the current calibration and

use the previous calibration.

6. Change

the

Automatic

calibration

level

or

Highlight the Auto

calibration level.

Table 1: Calibration Levels

2

Accuracy Microphone

Class

Directional

Response of

Microphone

Acoustic

Calibrator

Calibration

Level

3

Class 1

Instruments

CEL-250

Free Field (IEC)

CEL-110/1

114.0 dB

CEL-284/2

114.0 dB

Random Incidence (ANSI)

CEL-110/1

113.8 dB

CEL-284/2

113.8 dB

Class 2

Instruments

CEL-485

Free Field (IEC)

CEL-110/1

CEL-110/2

CEL-282

CEL-284/2

113.6 dB

Random Incidence (ANSI)

CEL-110/1

CEL-110/2

CEL-282

CEL-284/2

113.6 dB

Note 2: These levels may also be subject to further correction as follows:

(a)When a traceable calibration certificate is available for the acoustic calibrator
and/or the sound level meter.

(b)To compensate for atmospheric conditions as described in the calibrator
instructions.

Note 3: The instrument can be set to Autocalibrate to these levels.

Page 40 - CEL-450/490 Operators Handbook

Acoustic Calibration Check