Remove the batteries when the instrument is to be out of

service for some time. A built in lithium battery will maintain

stored data and setup information while the instrument is
switched off.
External DC Via 2.1 mm Power Connector: 9 to 14 VDC

at typically 150 mA with 1000 mA inrush current.

AC Output:

Approximately 0.5 VRMS via “AC Line Out” jack socket.
Full scale output corresponds to either 94 or 114 dB, with 22 k

W

minimum load impedance.
This output is suitable for DAT recorders, tape recorders, PC wav

file recording and headphone applications.

AC Input:

10 VRMS maximum via “AC Line In” jack socket or via a

CEL-516 Dummy Microphone. Maximum source impedance is
100 k

W.

This input is suitable for inserting direct-line signals from DAT or

tape recorders for analysis.

Optional DC Output:

0 to 2 VDC corresponds with 0 to 140 dB, with 2 k

W output

impedance.

Internal Clock:

Date and time accuracy better than 2 seconds per day.

Electromagnetic Compatibility:

The instrumentation is designed and tested to comply with the

following EMC and ESD Standards.
IEC 61000-4-2 Testing and Measuring Techniques - Electrostatic

Discharge Immunity Tests,

IEC 61000-4-3 Electromagnetic compatibility (EMC) - Radiated

Electromagnetic Field Tests.

IEC 61000-4-6 Electromagnetic compatibility (EMC) - Immunity to

Conducted Disturbances induced by Radio Frequency Fields.
Tested at 10 V/m or greater.