2 automatic calibration check of the microphone – Casella CEL CEL-450 User Manual

Page 38

Advertising
background image

DO NOT remove the protective metal grid from Class

1 microphones.

With a WS2, 1/2"

microphone, fit the

calibrator directly on to
the microphone, making
sure it is pushed firmly

into contact with the stop

in the calibrator cavity
(Figure 5).

With a WS3, 1/4"

microphone, fit the
coupler, supplied with

the calibrator, on to the

microphone making sure
it is pushed firmly into
contact with the stop in

the coupler cavity (Figure 5). Then fit the coupler complete with

microphone and instrument into the calibrator cavity, again ensuring that
it is pushed firmly into contact with the stop in the calibrator cavity.

DO NOT lay the sound level meter and calibrator on a

horizontal surface during calibration, as the combined
weight will cause the microphone to move inside the
calibrator cavity with the risk of causing damage and the
possibility of obtaining an incorrect calibration level.

Support the sound level meter and calibrator in an upright

position. To aid removal, the coupler flange does not fit tightly against

the calibrator housing.

A Calibration screen is displayed at the end of the start up

sequence and this will be the normal entry to the calibration check. The
Calibration screen can also be obtained via the Calibration option on the

Main menu. Perform an automatic calibration check of the microphone
as detailed in Section 4.2, or a manual check as described in Section 4.3

4.2 Automatic Calibration Check of the

Microphone

The Calibration screen shown after start up allows automatic calibration.

Figure 5: Fitting the acoustic calibrator

Page 38 - CEL-450/490 Operators Handbook

Acoustic Calibration Check