SPECIFICATION

6.1 General

Standards:

The CEL-450/490 series are self contained sound level meters
designed to comply with the following international standards.
IEC 61672-1 2002-5 (Electroacoustics - Sound Level Meters)
Group “X” instruments, performance class 1 or 2 as relevant to

the instrument model.
IEC 60651 197, IEC 60804: 1985, ANSI S1.4: 1983, ANSI S1.
In addition, for Octave and Third-Octave band versions (B and C

models) the filters comply with:
EN 61260: 1996, Class 0 and ANSI S1.11, Order-3 Type 0C.
Octave Filters: 11 bands with centre frequencies from

16 Hz to 16 kHz (B and C models).
Third Octave Filters: 33 bands with centre frequencies from

12.5 Hz to 20 kHz (C models).
Narrow band filters may also be pre-weighted with A, C or Z

frequency weightings.

Measurement Ranges:

Single 0 -140 dB range: Linear operating range defined by

the self-generated noise of the instrument only.
Peak levels: A, C and Z available to 143 dB.
Dynamic range: Typically 123 dB.
Level Detector: Digitally derived true root-mean-square (RMS)

detection, 0.1 dB display resolution.

Electric Noise Floor:

Noise floor with 18 pF dummy microphone: Typically

16.5 dBA.
Total inherent Noise including microphone thermal
noise at 20

o

C:Typically <18.5 dBA.