9.

Record about 30 seconds of the calibration signal.

A calibration signal with known level is inserted via the sound

level meter and recorded, so that when replayed, it gives an
indication of the record/replay characteristics of the tape.

10.

Switch the calibrator OFF and remove it from the microphone.

When the VU reading is estimated (which it usually must be at
these levels), repeatability from one series of recordings to

another may suffer. Therefore, either keep the Record Level

control in the same position for ALL recordings, or perform a
new calibration for EACH series of measurements.

11.

Without touching the Record Level on the recorder, if required,
change the measurement range of the instrument to

accommodate the noise signal.

12.

Make a note of the range used for measurement.

13.

Proceed with recording the noise signal.

5.5.2 Replay

The following guide lines are suggested for replaying sound signals.

Figure 6: Bottom panel of the instrument

Page 60 - CEL-450/490 Operators Handbook

Operation