9.

Use the filter set nudge buttons to position the arrow above
the identity number (2 to

6) of the memory location in which

data is to be stored.

If the display shows the "Err ALL" message, check the settings
of the sound level meter to ensure that the filter set is in
circuit, and that the meter is set to measure Leq values.


10. Set the CEL-296/3 BAND switch to "Oct" or 1/3 Oct" according

to the width required for the measuring procedure.

The filter set display will now Indicate "Err ALL".


11. Press the sound level meter PAUSE button to release the

pause condition when it is required to begin the autoscan.

The display changes to indicate sound level plus frequency.


Microprocessor Pre-start Checks

Autoscans start as soon as the sound level meter pause condition is
released, provided the filter set microprocessor is satisfied with the
results of the following checks.

a)

"FA" or "FL" has been selected on the sound level meter, so
that the filter set Is actually included in the circuit.


b)

Sound level meter is set to determine the energy averaged
sound level (Leq).


c)

Sound level meter time weighting Is set to either "S" or "F".

It is not recommended that autoscan be used for impulsive
noise analysis, which should be performed under manual con-
trol using memory 1 as the store location.


d)

Digital data returned from the meter in the "All Pass" band
must be at least 10 dB above the lowest measurable reading

CEL Instruments Ltd

CEL-296/3 Operator Handbook

Page 39