background image

3.15 kHz

10.0 dB

< 10.5 dB(A)

4 kHz

10.0 dB

< 10.5 dB(A)

5 kHz

10.0 d8

< 10.5 dB(A)

6.3 kHz

10.0 dB

< 10.5 dB(A)

8 kHz

10.0 dB

< 10.5 dB(A)

10 kHz

10.0 dB

< 10.5 dB(A)

12.5 kHz

10.1 dB

< 10.5 dB(A)

16 kHz

10.8 dB

< 10.5 dB(A)

20 kHz

11,8 dB

< 10.5 dB(A)

Octave bands:

Band

Linear

A-weighted

8Hz

24.2 dB

<11 d8 (A)

16 Hz

19.0 dB

< 11 dB(A)

31.5 Hz

16.0 dB

< 11 dB(A)

63 Hz

14.2 dB

< 11 dB(A)

125 Hz

12.6 dB

< 11 dB(A)

250 Hz

12.0 dB

< 11 dB(A)

500 Hz

10.3 dB

< 11 dB(A)

1 kHz

10.2 dB

< 11 dB(A)

2 kHz

10.3 dB

< 11 dB(A)

4 kHz

10.0 dB

< 11 dB(A)

8 kHz

11.3 dB

< 11 dB(A)

16 kHz

13.9 dB

< I 1 dB(A)


Effect of humidity on accuracy:

Less than ±0.5 dB over the range 30 to 90% RH (provided
there is no condensation). relative to the value at 65% RH and
40°C.

Page 60