Line direction functiondescription – Casella CEL CEL-296 User Manual

Page 55

Advertising
background image

Table 1 Connector lines used in top connector for interface

Line Direction Function

Description

2

--

Serial ground

0 V

9

Output

Request to send Line is set low by sound level

meter when data is available
for output, data is valid ONLY
while this line is low

10

Output

Serial data

Serial data output from sound
level meter

12

Input

Clear to send

Line must be set low by filter
set to enable output from
sound level meter.

Table 2 Connector lines used in bottom connector for interface

Line Direction Function

Description

2

--

Serial ground

0 V

3

Input

Filter step

Taken to +5 V for minimum of
25 ms then opened

6

Input

All stop reset

Closed to line 2 for minimum of
25 ms then reopened

9

Output

Request to send Line is set low by filter set

when data is available for out
put, data is valid ONLY while
this line is low

10

Output

Serial data

Serial data output from filter
set

11

Input

Clear to send

Line must be set low by
receiving device to enable
output from filter set

CEL Instruments Ltd