4 configure criterion level, Configuration – Casella CEL CEL-310 User Manual

Page 18

Advertising
background image

3.

Use

and

to select the required setting (see Table 1).

4.

Press

again to save the setting and return to the Main Menu.

3.4

Configure Criterion Level

The criterion level (CL) is the level that gives 100% dose in a standardised
8 hour day. This level is used in the calculation of L

EP.d

or TWA, %Noise

Dose and Estimated %Noise Dose. The ability to configure the Criterion
Level enables Dose Badge measurements to be made to suit local,
national or international standards. Settings to be used for some
International Standards are shown in Table 1.

PLEASE NOTE that changes to the Dose Badge configuration

are programmed into the Dose Badge ONLY when it is Reset.

1.

While displaying the Main Menu, use

or

to select

Criterion Level.

2.

Press

to display the Criterion Level menu.

3.

Use

and

to select the required setting (see Table 1).

4.

Press

again to save the setting and return to the Main Menu.

M-81

Select Criterion Time

12 hours
16 hours
18 hours

8 hours

Four options are shown, with the
current setting highlighted.

OK

M-09

Reader Main Menu

CL= 85 dB

Adjust LCD Contrast
Time History Control
Criterion Time

Criterion Level

The current Criterion Level setting
is shown in the message line

OK

M-91

Select Criterion Level

80 dB

87 dB
90 dB

85 dB

Four options are shown, with the
current setting highlighted.

OK

Configuration

Page 18 - CEL-310 Dose Badge Users Handbook