2 reset the dose badge, Operation – Casella CEL CEL-310 User Manual

Page 21

Advertising
background image

A Calibration ok message is displayed when calibration has
been completed satisfactorily.

If the calibration cannot be completed sucessfully, the Reader
may display one of the following error messages:

Too high
Too low
Badge running.

Refer to Chapter 6: Troubleshooting for assistance.

4.2

Reset the Dose Badge

All dose badges to be used for the measurement task must have their
data stores reset to zero, before new measurements can be made.

Install the Dose Badge in the calibrator cavity with infrared

windows oriented as described in Section 4.1 (Figure 5).

1.

If not already on, press

to switch the Reader ON.

2.

Wait for the status screen to be displayed, then press

to

initiate the Reset sequence.

C-02

Calibrated to 114 dB

Calibrating . . . . ok

Reset

C-03

Reset Noise Badge

OK

Cancel

Reset progress

C-04

Resetting . . . . ok

Operation

CEL-310 Dose Badge Users Handbook - Page 21

Advertising