7 print stored measurements – Casella CEL CEL-310 User Manual

Page 26

Advertising
background image

The Reader downloads the measurements before it downloads
Time History. If there are errors during the download of Time
History data, the Reader will store the measurements and show
Trace Error in the message line at the bottom of the display.

DO NOT REMOVE the Dose Badge until the Reader shows
Reading ok to indicate a successful download.

Once the measurement download has been completed, the
Reader will display the last measurement.

4.

Download data from the other Dose Badges used for the job in
the same way.

4.7

Print Stored Measurements

The currently displayed measurement can be printed by a serial printer
connected via the Reader unit’s RS232 Port. The communications
protocol used by the Reader is as follows:

Baud

38,400

Parity

None

Data Bits

8

Stop Bits

1

1.

Press

to display the Main Menu.

2.

To print the current measurements, use

to select Print

Current Data (see the fold-out diagram at the back of this
handbook).

3.

Press

to confirm the choice.

The data from the current measurement will be printed.

4.

To print all stored measurements, use

to select Print All

Data (see the menu diagram at the back of this handbook).

5.

Press

to confirm the choice.

The data from all measurements will be printed.

If there are no measurements in the memory, the Reader will
display a No Measurements message.